Belec Compact Port HLC – Technical Specifications

Belec Compact Port HLC - Technical Specifications

Optics

  • Spectrometer in Paschen-Runge mounting, using the latest 6GSO-System (6th Generation Spectrometer Optic System)
  • Rowland circle diameter 300 mm
  • Usable wavelengths 145-410 nm
  • Zeiss-Grating with 3600 lines/mm
  • Reciprocal dispersion 0.9 nm/mm (1st order)
  • Shock resistance
  • Temperature stabilized detectors with wavelength depending configured entrance window (US-licensed)
  • High-speed photomultipliers for perfect carbon analysis and determination of h-and l-alloyed steels
  • Integrated noise compensation

Sparking Probes

  • Argon-flushed sparking probe for exact analysis, including C, P, and S. N optional
  • Argon flow 0.1 l / min in standby and 2.5 l / min during analysis
  • Argon control on probe plug, independent from instrument
  • Air probe for quick mix-up checking
  • Adapters for wires, pipes and small parts available for all probes
  • Lightweight shockproof plastic probe housing
  • Start and clear buttons easily hand-operated
  • Signal on mix-up identification: visual display for “repeat” and “reject”, start button is blocked until confirm button is pressed
  • Multi fibre quartz optics, standard lengths 3m to 8m, optional
  • Low-wear tungsten electrode
  • Silver electrode for air probe
  • Probe connector system
  • Adapters available for different sample types as wires, tubes or small parts
  • Customized adapters on demand

Source

  • Sparking generator with maximum 400 Hz frequency
  • Unipolar discharge
  • Separate parameter for pre-sparking and integration selectable via software
  • Ignition frequency program specifically selectable via software
  • Discharge parameters specifically selectable via software
  • Arc source for air probe, optional
  • Ignition voltage 20 kV

Electronics

  • Temperature stabilization, at 0.1°C exact
  • Separate AD- converter board for each detector, mounted on multi channel board, coupled by high-speed USB port
  • Integrated noise suppression
  • integrated background compensation
  • Unlimited numbers of measuring channels, configurable for several bases

Dimensions

  • Width 16.5 in. (420 mm)
  • Height 7.9 in. (200 mm)
  • Depth 19.3 in. (490 mm)

Weight

  • Analysis unit 37.5 lbs. (17,00 kg)
  • Argon Probe 2.9 lbs. (1,30 kg)
  • Argon Probe UV 3.9 lbs (1,75 kg)
  • Air Probe 1.8 lbs. (0,80 kg)

* handheld weight

Power Supply

  • 230V/50Hz or 110V/60Hz
  • 100W in stand-by mode
  • 600W during analysis

Computer Hardware

  • System-integrated industrial computer system
  • Intel® Atom™ Processor N270 (1.60 GHz)
  • 1 GB RAM (max. 2 GB)
  • 2.5” hard disk 60 GB minimum
  • 10.4” TFT colour display
  • Touch screen
  • Special dust and moisture protected keyboard integrated in hinged lid
  • 3 x USB 2.0 ports
  • Ethernet interface

Belec WIN 21 Analysis and Quality Control Program


Software

  • Arbitrary operating system, e.g. Windows 7
  • Remote-Service-System
  • Display of analysis values at each measurement
  • As many analysis programs to customer specifications as required
  • Individual analysis parameters for each program
  • Automatic Program Finding (APF)
  • Analysis computation with: background correction, curve position correction, additive and multiplicative inter-element correction
  • Automatic correction with standard types
  • Easy and simultaneous recalibration of several programs
  • Extended recalibration cycles by usage of MCDC (Multi Channel Drift Correction)
  • Mix-up checking by comparison with reference measurement
  • Grade Test by comparison with grade ranges
  • Type calibration and type measurement
  • Tolerances for every analysis program and element in absolute and relative weight percentages, individually adjustable
  • Average and standard deviation from chosen measurements
  • Warning signal, when calibration curve is exceeded
  • Automatic reminder of regular recalibration
  • Automatic display of grade name or material number
  • Alloy data bank, 100.000 grades and more storable (only limited by computer storage capacity)
  • Text size on monitor variable for optimum legibility
  • Protocol storage function
  • Report memory function for later analysis, printing and archiving
  • Several statistic functions with graphical representation